Technical Specifications
NESS L300 Foot Drop System Control Unit Specifications
Classification Internally powered, continuous operation.
Operation Modes Gait, Training, Clinician and Standby.
Battery type Rechargeable AAA NiMH 1.2 V, 900 mAh.
Controls On/Off illuminated button.
Gait Mode illuminated button for changing operation Modes.
Intensity +/- buttons to fine-tune the intensity.
Volume +/- buttons control buzzer volume.
Indications Three status LEDs: Control Unit, RF Stim Unit and Intelli-Sense Gait Sensor.
Numerical display designates relative stimulation intensity.
Illuminated buttons designate system operation Mode.
"Beeps" for audible indications.
Carrying In pocket, neck strap, wrist strap or belt pouch.
Dimensions Height: 71 mm (2.8").
Width: 46 mm (1.8").
Depth: 17.5 mm (0.7").
Weight 45 grams (1.5 oz).
Environmental  Transport and storage temperature: -20° to 60°C (-4 to 140°F).
Operating conditions temperature: 5° to 40°C (41 to 104°F).
Charging temperature: 5° to 40°C (41 to 104°F).
Relative humidity: 25% to 85%.
Atmospheric pressure: 900 hPa to 1060 hPa.
 
NESS L300 Foot Drop System RF Stim Unit Specifications
Classification Internally powered, continuous operation with type BF applied parts.
Operating voltage 3.7 V.
Battery type Proprietary rechargeable LiIon (Lithium Ion) 3.7 V, 700 mAh.
Indications Status (fault, battery, charging) and Stimulation LEDs "Beeps" for audible indications
Dimensions Height: 74 mm (2.9").
Width: 43 mm (1.7").
Depth: 15 mm (0.6").
Weight 50 grams (1.6 oz).
Environmental Transport and storage temperature: -20° to 60°C (-4 to 140°F).
Operating conditions temperature: 5° to 40°C (41 to 104°F).
Charging temperature: 5° to 40°C (41 to 104°F).
Relative humidity: 25% to 85%.
Atmospheric pressure: 900 hPa to 1060 hPa.
 
NESS L300 Foot Drop System Pulse Parameters
Pulse Balanced Biphasic.
Waveform Symmetric or Asymmetric.
Intensity
0-90mA [0-80mA in 1 mA resolution (positive phase)].
Max Voltage 120 V .
 
Symmetric
Asymmetric
 
Positive Pulse duration 100 µs 200 µs 300 µs   100 µs 200 µs 300 µs  
Negative Pulse duration 100 µs 200 µs 300 µs   400 µs 800 µs 1200 µs  
Inter-phase interval
50 µs
0
 
Total Pulse duration 250 µs 450 µs 650 µs   500 µs 1000 µs 1500 µs  
Max Load 5000 Ω (Subject to Max Voltage limitation).
Pulse repetition rate 20-45 Hz (5 Hz resolution).
Gait Parameters
Ramp up 0 - 2 seconds, 0.1 second resolution.
Ramp down 0 - 2 seconds, 0.1 second resolution.
Extend (Delay) 0 - 100% of stance time, 10% resolution.
Max duration 4 seconds.
Training Parameters
On time 4 - 20 seconds, 1 second resolution.
Off time 4 - 60 seconds, 1 second resolution.
Ramp up 0 - 2 seconds, 1 second resolution.
Ramp down 0 - 2 seconds, 1 second resolution.
Total time 1 - 60 minutes.
 
NESS L300 Foot Drop System Leg Cuff Specifications
Leg Cuff Fabric-polymer Leg Cuff.
Limb Size 29 – 51 cm (11" – 20") limb circumference.
Dimensions Height: 135 mm (5.3").
Width: 100 mm (3.9").
Depth: 125 mm (4.9").
Weight Approximately 150 grams (4.8 oz).
 
NESS L300 Foot Drop System Electrode Specifications
Electrodes Two, 45 mm hydrogel Electrodes.
Note: Use only Electrodes provided by NESS Ltd. or
Bioness.
Electrode Base Two relocateable polymer Electrode Bases for individual fitting.
 
NESS L300 Foot Drop System Intelli-Sense Gait Sensor Specifications
Classification Internally powered, continuous operation with type BF applied
part(s).
Battery type Lithium coin cell, CR2430, 280 mAh.
Dimensions of
the transmitter
Height: 80 mm (3.2").
Width: 50 mm (2.0").
Depth: 10 mm (0.4").
Weight 35 grams (1.1 oz).
Environmental Transport and storage temperature: -20° to 60°C (-4 to 140°F).
Operating conditions temperature: 5° to 40°C (41 to 104°F).
Relative humidity: 25% to 85%.
Atmospheric pressure: 900 hPa to 1060 hPa.
 
NESS L300 Foot Drop System Power Supply Specifications
Use medical Class II safety approved power supply provided/approved
by NESS Ltd. or Bioness with the following ratings:
Input
Voltage 100-240V AC.
Current 200 mA.
Frequency 50-60 Hz.
Output
Voltage 5V ± 5%.
Current 1300 mA.
Current Note: Do not use the Control Unit or RF Stim Unit while charging.
 
NESS L300 Foot Drop System Wireless Link Specifications
Frequency band 2.4 GHz, ISM band.
Transmission power Complies with FCC 15.247 (for US) / ETSI EN300-440 (for  
Europe) Regulations.


© 2008 Bioness Contact Us | Careers | Site Map | Privacy Policy | Terms of Use

NESS®, NESS L300™, NESS H200™, Intelli-Gait™ and Intelli-Sense Gait Sensor™ are trademarks of N.E.S.S. Neuromuscular Electrical Stimulation Systems, Ltd., Ra'anana, Israel | Manufacturer: N.E.S.S. Neuromuscular Electrical Stimulation Systems, Ltd.| North American Distributor: Bioness Inc. | Bioness®, the Bioness Logo®, and LiveOn™ are trademarks of Bioness, Inc., Valencia, CA | The NESS H200 and the NESS L300 are RX Only