Text Size

> Bioness for Hand Rehab
Bioness for Foot Drop
Success Stories
About Bioness
Getting Started
Contact Us

Print | Email
 
Home > Bioness for Hand Rehab > Technical Specs
Technical Specifications
NESS H200 Hand Rehabilitation System Technical Specifications
Type of Electrical Stimulation Constant voltage, biphasic symmetrical pulse.
Microprocessor-controlled automatic operation Single current source multiplexed among five surface Electrodes on the Orthosis.
Dimensions and weight Control Unit (incl. Batteries)
3.6 x 1.6 x 6.8 (in.), 0.83 lb.
9.1 x 4.0 x 17.2 (cm), 377 gm.
Orthosis Medium and Small
3.5 x 3.0 x 10.8 (in.), 0.72 lb.
8.9 x 7.7 x 27.5 (cm), 325 gm.
Orthosis Large
4.2 x 3.7 x 12.0 (in), 0.87 lb.
10.7 x 9.5 x 30.5 (cm), 396 gm.
Battery 9.6 V Rechargeable Nickel-Metal-Hydride.
Holds 280mAH charge, giving 15 hours continuous operation at typical power.
Charger AC Adapter, Class 2 Transformer,
Input: 100-240 VAC, 50-60 Hz.
Battery Indicator Charge Indicator.
Battery low visual display and audio beep warning.
10 selectable stimulation
intensity levels
Push-button adjustment of pulse intensity.
Ramp up - ramp down in all Modes.
7 selectable operating Modes


Carrier wave
Personal (pre-set program or personalized), ExerciseF, Open-ExerciseF, Grasp-ExerciseF, OpenF, Grasp & Key.
Wave form: Sinusoidal, balanced, no DC component.
Frequency: 11 kHz.
F=Fast speeds available
Stimulation pulse burst
Stimulation Pulse Burst graph for the NESS H200 Hand Rehabilitation System Technical Specifications

Personal, Exercise, Open-Exercise, Grasp-Exercise and Open Modes:
Duration: 10 levels from 0.01 mS to 0.35mS.
Frequency: 36 Hz.
Grasp and Key Modes:
Duration: 10 levels from 0.01 mS to 0.5 mS.
Frequency: 18 or 36 Hz.
Pulse Voltage and Current (all Modes):
115V rms; 58 mA rms, at 2kΩ load.
75V rms; 150 mA rms, at 500Ω load.
Average parameters Maximum voltage:
1.46V rms at 2kΩ load (Exercise Mode).
Maximum current:
1.9 mA rms at 500Ω load (Exercise Mode).
Maximum power 0.13 Watts at 500Ω load (Exercise mode).
Automatic shutdown Exercise, Open-Exercise and Grasp-Exercise Modes: 90 minutes.
Open, Grasp, and Key Modes: 45 minutes.
Personal Mode: programmable up to 99 minutes.
Classification Internally powered, Type BF, Continuous operation.
Transport and storage -20° C to 60° C (-4°F to 140°F) ambient temp.
10% to 85% relative humidity.
500hPa to 1060hPa atmospheric pressure.
Operating 5°C to 40°C (40°F to 105°F) ambient temp.
10% to 85% relative humidity.


© 2008 Bioness Contact Us | Careers | Site Map | Privacy Policy | Terms of Use

NESS®, NESS L300™, NESS H200™, Intelli-Gait™ and Intelli-Sense Gait Sensor™ are trademarks of N.E.S.S. Neuromuscular Electrical Stimulation Systems, Ltd., Ra'anana, Israel | Manufacturer: N.E.S.S. Neuromuscular Electrical Stimulation Systems, Ltd.| North American Distributor: Bioness Inc. | Bioness®, the Bioness Logo®, and LiveOn™ are trademarks of Bioness, Inc., Valencia, CA | The NESS H200 and the NESS L300 are RX Only